Assessment of endothelial function and myocardial flow reserve using 15O-water PET without attenuation correction.
분류 PET 조회 2280
발행년도 2015 등록일 2015-10-17
출처 Eur J Nucl Med Mol Imaging (바로가기)
PURPOSE:
 
Myocardial blood flow (MBF) measurement using positron emission tomography (PET) from the washout rate of 15O-water is theoretically independent of tissue attenuation. The aim of this study was to evaluate the impact of not using attenuation correction in the assessment of coronary endothelial function and myocardial flow reserve (MFR) using 15O-water PET.
 
METHODS:
 
We retrospectively processed 70 consecutive 15O-water PET examinations obtained at rest and during cold pressor testing (CPT) in patients with dilated cardiomyopathy (n = 58), or at rest and during adenosine infusion in heart transplant recipients (n = 12). Data were reconstructed with attenuation correction (AC) and without attenuation correction (NAC) using filtered backprojection, and MBF was quantified using a single compartmental model. The agreement between AC and NAC data was assessed using Lin's concordance correlation coefficient followed by Bland-Altman plot analysis.
 
 
(후략)