Brain PET imaging optimization with time of flight and point spread function modelling. | |||
---|---|---|---|
분류 | PET | 조회 | 2302 |
발행년도 | 2015 | 등록일 | 2015-10-17 |
출처 | Phys Med (바로가기) | ||
PURPOSE:
To assess the influence of reconstruction algorithms and parameters on the PET image quality of brain phantoms in order to optimize reconstruction for clinical PET brain studies in a new generation PET/CT.
METHODS:
The 3D Hoffman phantom that simulates 18F-fluorodeoxyglucose (FDG) distribution was imaged in a Siemens Biograph mCT TrueV PET/CT with Time of Flight (TOF) and Point Spread Function (PSF) modelling. Contrast-to-Noise Ratio (CNR), contrast and noise were studied for different reconstruction models: OSEM, OSEM + TOF, OSEM + PSF and OSEM + PSF + TOF. The 2D multi-compartment Hoffman phantom was filled to simulate 4 different tracers' spatial distribution: FDG, 11C-flumazenil (FMZ), 11C-Methionine (MET) and 6-18F-fluoro-l-dopa (FDOPA). The best algorithm for each tracer was selected by visual inspection. The maximization of CNR determined the optimal parameters for each reconstruction.
(후략)
|
|